Performance Specifications
Controlled Range Temperature 5-55 °C¹
Humidity @23°C 5 to 95 % RH²
Humidity @30°C 5 to 80 % RH²
Humidity @35°C 5 to 70 % RH²
Humidity @40°C 5 to 60 % RH²
Humidity @50°C 5 to 50 % RH²
Total Unit Uncertainty Temperature ± 0.1 °C
Humidity ±1.0 % RH
Uniformity Temperature 0.10 °C
Humidity 0.30 % RH @ 23 °C
Stability Temperature 0.05 °C
Humidity 0.15 % RH @ 23 °C
Ramp / Soak Rate of Change Temperature High → Low 1.5 °C / minute (Typical)
Low → High 5.0 °C / minute (Typical)
Humidity High → Low 5.0 % RH / minute (Typical)
Low → High 10 % RH / minute (Typical)
Warm-Up Period With Recent Use Twice the time since last warmed up (Max 30 Minutes)
Cold 30 Minutes
General Specifications
Calibration Traceability 17025 Traceable to NIST
Accrediting Body NVLAP
Purchasing Charged Separately
Operating Ambient Conditions Temperature 18 °C to 28 °C
Humidity Up to 80 % RH
Storage Conditions Temperature -20 °C to 50 °C
Humidity 0 % to 95 % RH (non-condensing)
Altitude 2000 m
Calibration Traceable to NIST
Resolution Display Temperature 0.1 °C
Humidity 0.1 % RH
Dew Point 0.1 °C (Calculated)
USB Temperature 0.1 °C
Humidity 0.1 % RH
Consumables
Water Reservoir 350 mL
Spill Resistant Yes
Required Fluid Distilled Water
Est Refill Period 1 Month (Typical)
Fill Indicator Floating Ball
Desiccant Type Molecular Sieve
Replacement When Indicating Desiccant is 3/4 Used
Replacement Frequency Depends Entirely on User Workload
Recalibration Frequency Depends on User Uncertainty Requirements
Once Per Year Recommended
Electrical
AC Mains Voltage Universal 100 - 260 V AC - 50/60 Hz
Standard Frequency Range 47 Hz to 63 Hz
Power consumption 300 VA
Mains Fuse Rating F4A 250 V (fast blow)
Transducer Power Output Voltage 12 V DC
Amperage 1 A Maximum
Fuse F 1A 250V (fast blow)
Mechanical
Generator Dimensions Measurements Width Depth Height
Metric 46.00 cm 38.50 cm 23.50 cm
Imperial 18.11 in 15.16 in 9.25 in
Chamber Dimensions Measurements Diameter Depth
Metric 14.50 cm 20.00 cm
Imperial 5.71 in 7.87 in
Working Dimensions and Volume Measurements Diameter Depth
Metric 12.50 cm 10.90 cm
Imperial 4.92 in 4.29 in
Volume 2 Liter Effective Working Volume
Weight Unit Only Metric 12.0 kg
Imperial 26.5 lb
Unit With Case Metric 29.2 kg
Imperial 64.4 lb
Probe Power Supply Power Supply 12 Volt DC @ 1A
Standard Port Quantity 6 Ports
Standard Port Sizes Imperial 0.000 in (Plug)
0.125 in
0.250 in
0.375 in
0.500 in
0.625 in
0.750 in
0.875 in
Metric 0.000 mm (Plug)
3.175 mm
6.350 mm
9.525 mm
12.700 mm
15.875 mm
19.050 mm
22.225 mm